Products Introduction

Content Introduction
The NI PXI-4472 is a legacy 8-channel dynamic signal acquisition module designed for sound, vibration and acoustic measurement in PXI automated test systems. It adopts 24-bit delta-sigma ADC architecture with a maximum sampling rate up to 102.4 kS/s, supporting built-in IEPE excitation for accelerometers and microphones. It features automatic adjustable anti-aliasing filters that adapt to configured sampling rates, and software-selectable AC/DC coupling with a 3.4 Hz high-pass cutoff for AC mode. Multiple modules can be synchronized via RTSI bus and PXI trigger bus to expand channel count for large-scale structural testing. Widely used in automotive NVH testing, machinery condition monitoring and acoustic validation benches, this module is commonly stocked as spare replacement hardware.

Product Details
The NI PXI-4472 is a single-slot 3U PXI DSA module with eight simultaneously sampled analog input channels, no analog output channels.
It provides integrated IEPE constant current excitation to power IEPE accelerometers and condenser microphones directly.
AC or DC coupling is configurable for each channel, the AC coupling high-pass filter cutoff is 3.4 Hz.
Anti-aliasing filters automatically adjust according to the set sampling rate to simplify measurement configuration.
Front-panel BNC connectors provide stable connection for vibration transducers, microphones and standard test cables.
Multi-module synchronization is supported through the RTSI bus and PXI trigger bus for high-channel measurement systems.
It runs under NI-DAQmx driver and supports programming via LabVIEW, LabWindows/CVI, C/C++, Python and IVI compliant software.
This is a discontinued National Instruments dynamic signal acquisition module.
Typical applications include multi-channel vibration monitoring, noise testing, structural modal testing and industrial equipment fault diagnosis.

Product Parameters
Part Number: PXI-4472 (778279-01)
Brand: National Instruments (NI)
Form Factor: 3U PXI Hybrid, single slot
ADC Resolution: 24-bit delta-sigma
Analog Input Channels: 8 simultaneously sampled channels
Max Sampling Rate: 102.4 kS/s
Dynamic Range: 110 dB typical
Coupling Mode: Software selectable AC / DC coupling
AC High-pass Cutoff: 3.4 Hz
IEPE Excitation: Built-in support for IEPE sensors
Anti-Alias Filter: Automatically adjusted anti-aliasing filter
Connector Interface: 8 × BNC female
Synchronization: PXI trigger bus and RTSI bus support
Driver: NI-DAQmx
Compatible Software: LabVIEW, LabWindows/CVI, C/C++, Python, .NET, IVI
Operating Temperature: 0 °C to 55 °C

Tags: NI PXI-4472 PXI-4472 DSA 8-channel vibration module 24-bit sound acquisition